丙酮氣體檢測管102SA ,丙酮氣體檢測管102SA ACETONE CH3COCH3 0.1-5.0%,日本光明理化工業(yè)株式會社從1947年開始,在氣體分析領(lǐng)域開拓了六十余年 , 形成了于世的 ” 北川 ” 式氣體檢測管。氣體快速分析法 ” 北川 ” 的歷史就是日本檢測管的歷史 。目前是世界上zui大的檢測管生產(chǎn)廠家,也是日本*家通過美國SEI的認證,歐盟的認證的檢測管廠家。公司不僅有ISO9001:2000的認證,還獲得了環(huán)境管理的ISO14001的認證。北川式手動氣體采集器AP-20 備有采氣完成指示計 , 可以清楚的判明采氣是否完成 . • 有專門截留玻璃的截留槽 , 以防止檢測管割尖時玻璃片飛濺 . • 有采氣容積為 50 毫升和 100 毫升的兩段式制動卡 . • 采氣器極為小巧 , 并有便于攜帶的皮具包 。
日本光明 丙酮氣體檢測管102SA
Tube 02SA
1. PERFORMANCE
1) Measuring range :1.0-5.0% 0.1-2.0%
Number of pump strokes 1/2(50mr) 1(100mr)
2) Sampling time :1.5 minute/1 pump stroke
3) Detectable limit :200 ppm
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Orange→brown
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH3COCH3+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
0.1
0.2
0.5
1.0
1.5
2.0
%
GAS To pump
ACETONE
*Methanol is indicated with half the sensitivity and Ethyl acetate has the same sensitivity with Acetone.
(NOTE)
1) If the discolouration is over the scale, replace the tube with new one and pull the handle at half stroke(to 50mr
line). And read a figure from the scale on the tube.
2) Correct the reading value with theTEMPERATURE CORRECTION TABLE first . and convert the value into an
actual concentration by uing the conversion scale shown in the instruction sheet.
Substance Interference Coexistence
Alcohols FIG.1 Similar stain is produced. Higher readings are given.
Other ketones 〃〃
Aromatic hydrocarbons 〃〃
Esters FIG.2 〃〃
Halogenated hydrocarbons Wdishcoolelo rueraegde. nt is slightly 0.5% 〃
0
0.5
1.0
1.5
Methanol〔%〕?
FIG.1 Influence of Methanol
Acetone indicating value 〔%〕
0.5 1.0 0
0.5
1.0
1.5
Ethyl acetate〔%〕?
FIG.2 Influence of Ethyl acetate
Acetone indicating value 〔%〕
0.5 1.0
TEMPERATURE CORRECTION TABLE
Scale True Concentration (%)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(%) (32°F) (50°F) (68°F) (86°F) (104°F)
2.0 - 2.38 2.00 1.78 1.60
1.5 2.20 1.76 1.50 1.30 1.16
1.0 1.44 1.18 1.00 0.86 0.76
0.5 0.72 0.60 0.50 0.42 0.36
0.1 0.16 0.12 0.10 0.08 0.08
0.1
0.2
0.5
1.0
1.5
2.0 5.0
4.0
3.0
2.0
1.0
Conversion
scale (I)
(50mR actual value)?
(100mR concentration scale)?
(%)?
102SA